Analytical Technology A15 Specifications Page 37

  • Download
  • Add to my manuals
  • Print
  • Page
    / 41
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 36
ATI A15/66 Residual Sulfite Monitor
34
O&M Manual
Rev-W (8/10)
Figure 11 - Pump Head Assembly Exploded View
AIR FILTER REPLACEMENT
The air pump that supplies stripping air to the reactor assembly contains a small inlet filter. Over
time, this filter will accumulate dirt and dust removed from the air used by the pump. This filter should be
replaced once a year. Open the rear compartment of the chemistry module to access the pump. On the
inlet of the air pump, you will see a porous plastic filter connected with a short piece of silicon tubing.
Remove the old filter and discard. Replace with a new filter element. One spare filter is supplied with the
unit, and additional filters are listed on the spare parts list.
Page view 36
1 2 ... 32 33 34 35 36 37 38 39 40 41

Comments to this Manuals

No comments